Fluidics functions test – Bayer HealthCare Rapidlab 800 User Manual
Page 342

4Ć56
800 Series Operator's Manual
Test
Function
Roll Printer
Tests the ability of the roll printer to print all characters.
Bar Code Scanner
Tests the ability of the bar code scanner to read a test pattern.
Fluidics Functions Test
Use this procedure to test the ability of the reagent components to deliver reagents
to the fluid detectors in the reagent manifold and the measurement module.
1.
Access the Fluidics Functions Test screen from the Menu screen:
a. Select
3 Troubleshooting
and press
Enter
.
b. Select
1 Fluidics System
and press
Enter
.
c. Select
1 Fluidics Functions
and press
Enter
.
The Fluidics Functions Test screen appears as shown in Figure 4-2 for an 860
system.
Figure 4Ć2.ā Fluidics Functions Test Screen
Test one reagent
at a time.
Check here for results.
2.
Select the reagent you want to test and press
Enter
.
3.
Press
Start Test
.
4.
Check the screen for the message, Fluidics functions acceptable, to verify that
the system delivers the reagent to the fluid detectors.