Printing or transmitting qc sample reports, Figure 2ć28.ā qc sample report, Press – Bayer HealthCare Rapidlab 800 User Manual
Page 147

2Ć49
Operating the System
Printing or Transmitting QC Sample Reports
Use this procedure to print or transmit data from the QC Data Search Criteria Log
screen or the QC Data Search Result screen.
1.
Locate the sample you want as described in Recalling QC Data, page 2-45.
2.
Press
Reporting Options
.
If you want to . . .
Then . . .
print a QC sample report
a. Select Print QC Sample Report and
press
Enter
.
b. Press
OK
. Refer to Figure 2-28 for an example of
a QC sample report.
transmit a QC sample
report to an LIS or data
management system
a. Select Send Results and press
Enter
.
b. Press
OK
.
Figure 2Ć28.ā QC Sample Report
. . . . . . .
...........
..................... . . . . . . . . . . . . . . . . . . .
QC SAMPLE APR 12 1994 13:02
SYSTEM 850-1001 QC ID 123
Sequence No 00099 Level 1
Operator ID 12345 Lot 12345
Analysis Date Apr 12 1994 Expiration Dec 31 1995
Analysis Time 11:42
Results Units Target Range
pH 7.613 (7.594 - 7.634)
pCO
2
21.0 mmHg (19.8 - 23.8)
pO
2
148.8 mmHg (142.9 - 162.9)
Na+ 155.9 mmol/L (150.6 - 160.6)
K+ 7.0 mmol/L (6.72 - 7.72)
Ca++ 0.65 mg/dL (0.55 - 0.75)
Cl- 100 mmol/L (95 - 105)
Glucose 41 mg/dL (30 - 50)
Lactate 12.15 mmol/L (9.82 - 13.82)
ctHb 14.4 g/dL (13.5 - 15.5)
FO
2
Hb 1.3 % (0.7 - 1.7)
FCOHb 96.0 % (93.5 - 97.5)
FMetHb 0.5 % (0.2 - 0.6)
FHHb 2.2 % (2.1 - 3.1)
Accepted into QC File 1
QC File Information
File Status
840
850
........................... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
860
These results
appear when an
845, 855, or 865
is interfaced.
..............
.............. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CO-ox
3.
Press
Home
to return to the Ready screen.
The following result flags can appear on printed reports:
This flag . . .
Indicates . . .
*
the measurement did not reach endpoint.
↑
the result is above the upper limit of the target range.
(Continued)