System components, Measurement module components – Bayer HealthCare Rapidlab 800 User Manual
Page 37
1Ć15
Learning About the System
System Components
The 800 system components consist of the following functional groups:
S base model measurement module
S CO-ox model measurement module
S fluidic components
S electronic components
Measurement Module Components
The following illustrations show the measurement module components. Table 1-1
describes these components and their functions.
Table 1Ć1.āăMeasurement Module Components
Component
Description
Preheater
The preheater warms the sample to 37
°
C.
Sensors
The sensors detect analytes present in the sample and form the
actual sample path. The sample leaves the sample path through
the measurement module tubing after analysis. Refer to
Appendix I, Operating Principles, for more information about
each sensor.
Sample Ground/
Temperature Sensor
The sample ground/temperature sensor provides an electrical
grounding mechanism for stable sensor readings and also
measures the sample temperature.
Measurement Block
The temperature-controlled measurement block contains the
sensors and ensures a constant temperature of 37
°
C during
analysis.
Spring-loaded Latch
The spring-loaded latch keeps the sensors aligned in the
measurement block. The latch locks open for easy removal of
the sensors. The release tab on the latch releases the latch to
close it.
Contact Assembly
The contact assembly provides electrical contacts between the
sensors and the system.