Bayer HealthCare Rapidlab 800 User Manual
Page 541

Performance Characteristics
EĆ15
Level
Recovery
Expected
Observed
WRSD
*
n
Mode
Capillary
18
0.68
71.7
71.3
100.6
Microcapillary
24
2.24
71.0
71.3
99.5
Microsyringe
24
0.33
71.4
71.3
100.1
Expired Gas
24
0.17
71.1
71.3
99.7
* WRSD = within-run standard deviation
Table EĆ13.āă840 Recovery and Precision TestingĊpO
2
Level
Mode
n
WRSD
*
Observed
Expected
Recovery
1
Syringe
24
0.21
28.2
28.5
99.1
Capillary
18
0.40
28.3
28.5
99.3
Microcapillary
24
1.15
29.0
28.5
101.7
Microsyringe
24
0.17
28.6
28.5
100.3
Expired Gas
24
0.15
28.2
28.5
98.8
2
Syringe
24
0.19
50.0
49.9
100.3
Capillary
18
0.23
50.5
49.9
101.2
Microcapillary
24
0.71
49.0
49.9
98.3
Microsyringe
24
0.15
49.7
49.9
99.6
Expired Gas
24
0.14
49.7
49.9
99.7
3
Syringe
24
0.66
86.6
85.6
101.2
Capillary
18
0.36
84.8
85.6
99.1
Microcapillary
23
1.11
86.2
85.6
100.7
Microsyringe
24
1.08
85.8
85.6
100.3
Expired Gas
24
0.15
85.5
85.6
99.9
(Continued)