Figure 2ć36.ā twoćpoint calibration report, Press, When you finish – Bayer HealthCare Rapidlab 800 User Manual
Page 162: Done, Select next record and press, Select previous record and press, Select search criteria screen and press

2Ć64
Rapidlab 800 Operator's Manual
Figure 2Ć36.ā TwoĆPoint Calibration Report
2-POINT CALIBBRATION REPORT APR 12 1994 13:02
SYSTEM 860-1001
Sequence no 433
Calibration Date APR 12 1994
Calibration Time 11:42
CAL POINT
New Meas Drift Units
pH 7.382 7.381 -0.001
pCO
2
35.4 34.9 -0.5 mmHg
pO
2
85.0 85.0 0.0 mmHg
Na+ 140.0 140.9 0.9 mmol/L
K+ 4.00 4.00 0.00 mmol/L
Ca++ 1.25 1.26 0.01 mmol/L
Cl- 100 99 -1 mmol/L
Glucose 0 0 0 mg/dL
Lactate 0.00 0.02 0 mmol/L
pAtm 755 mmHg
SLOPE POINT
New Meas Drift Units
pH 6.838 6.839 0.001
pCO
2
70.8 71.3 0.5 mmHg
pO
2
0.0 -0.1 -0.1 mmHg
Na+ 100.0 100.3 0.3 mmol/L
K+ 8.00 8.04 0.04 mmol/L
Ca++ 2.50 2.52 0.02 mmol/L
Cl- 70 70 0.00 mmol/L
Glucose 180 180 0 mg/dL
Lactate 0.00 0.02 0 mmol/L
Atmospheric pressure
appears on all systems
. . . . . . .
...........
....................... . . . . . . . . . . . . . . . . . . .
840
850
............................. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
860
. . . . . . .
...........
........................... . . . . . . . . . . . . . . . . . . .
840
850
............................... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
860
7.
Press
Done
when you finish.
If . . .
Then . . .
more than one report is
found
the Done Options message box appears:
S
Select Next Record and press
OK
to view the next
report that appears on the log.
S
Select Previous Record and press
OK
to view the
previous report that appears on the log.
S
Select Search Criteria Screen and press
OK
to view the
Calibration Data Search Criteria screen.
S
Press
Cancel
to close this message box. The Search
Results screen appears.
one sample is found
the Calibration Data Search Criteria screen appears.
If the selected sample has a D code or status message associated with it, the View
Diagnostics F-key appears. Press View Diagnostics to view the Diagnostics
message box.
Procedural
Notes