Recalling the maintenance log – Bayer HealthCare Rapidlab 800 User Manual
Page 175

3Ć7
Maintaining the System
Recalling the Maintenance Log
Use this procedure to view and print the list of completed maintenance tasks for
the current or the previous month.
1.
Access the Maintenance Log screen from the Menu screen:
a. Press
4 Data Recall
and press
Enter
.
b. Press
3 Maintenance Log
and press
Enter
.
The Maintenance Log screen appears as shown in Figure 3-2.
Figure 3Ć2.āăMaintenance Log Screen
2.
Use the arrow keys to scroll through the task list.
3.
Press
Previous Month
to view tasks completed in the previous month.
4.
Press
Done
to return to the Menu screen.
The system only keeps two months of data online. At Autoclean time on the first
day of each month, the data is rotated and the oldest data is erased.
4
Menu Code
3