Preparing the 800 coćox module – Bayer HealthCare Rapidlab 800 User Manual
Page 608

HĆ16
800 Series Operator's Manual
Figure HĆ5.āăThe Rear Panels
Base Model
Interface
Network Port
Keyboard Port
Parallel Port
Bar Code Scanner
Port
Serial Port 1
Serial Port 2
Serial Port 3
3.
Open the 800 base model, secure the spring locks, access and remove the right
side panel by loosening the captive screws mounted on the interior floor of the
chassis. Do not remove the screws. Retain the side panel.
NOTE:ĄĂ
Slide the base model to the edge of the work surface to access the
front plug.
4.
Remove the two plugs mounted to the right side wall of the base model
hydraulic wall:
a. Push on the plug stems that protrude from the interior of the hydraulic wall.
b. Remove the plugs and retain.
Preparing the 800 COĆox Module
1.
Lift the hydraulic wall of the CO-ox module to access the power cable.
2.
From inside the CO-ox chassis push the connector end of the cable through the
slot and gently pull the cable through.