D5 no endpoint – Bayer HealthCare Rapidlab 800 User Manual
Page 310

4Ć26
800 Series Operator's Manual
D5 No Endpoint:
Glu
Lac
Problem
Sensor does not reach stable reading during predefined time limit.
Qualifiers:
Glu
Lac
Solutions
NOTE:ĄĂ
If D codes also exist for reagent problems such as D23, D24, D29, or
D50, correct these problems first, then correct the D5.
Perform a metabolite two-point calibration.
Perform the Measurement Test for the affected biosensor as described in
Measurement Test, page 4-70.
If the glucose or lactate biosensors have been on the system less than one day,
then perform three two-point calibrations.
Replace the biosensor as described in Replacing the Glucose and Lactate
Biosensors in Section 3.
If the D code reappears, contact your Service Representative.
D5