Bayer HealthCare Rapidlab 800 User Manual
Page 490

800 Series Operator's Manual
5Ć80
Figure 5Ć42.ā Patient Data Search Criteria Screen
This field displays the
number of samples
stored in system.
Enter a series of sample
sequence numbers for the
files you want to copy.
Enter a range of
dates for the files
you want to copy.
Use this screen to define the criteria for the patient data files you want to copy.
The system locates only the files that meet the criteria you specify. For
example, if you want to copy patient data files from a specific period, type in
the start date in the Analysis Date From field and the end date in the Analysis
Date To field. You can also copy files with consecutive sample sequence
numbers.
NOTE:ĄĂ
The system can copy up to 1500 patient samples on a single diskette.
If your search criteria produce too many samples to fit on a single diskette, the
system prompts you to reenter the search criteria to select fewer samples.
3.
Type the search criteria and press
Enter
after you complete each field. Press
Done
when you finish.
The Copying Files screen appears, with the Begin Copying Files dialog box.
4.
Insert a clean, formatted diskette in the diskette drive and press
Continue
.
The Writing to diskette message box appears while the system copies the files.
5.
When the Copy Files Finished screen appears, remove the diskette and label it.
If you do not insert a diskette, the No Diskette in Diskette Drive message box
appears. If you insert an unformatted or faulty diskette or a diskette that
already contains files, the Cannot Write to Diskette message box appears.
If . . .
Then . . .
you want to copy files
insert a clean, formatted diskette and press
Continue
.
you do not want to copy
files
press
Cancel
. The Copying Files screen appears,
with the Cancel Copy Files dialog box.
Procedural