Bayer HealthCare Rapidlab 800 User Manual
Page 539
Performance Characteristics
EĆ13
Table E-11 through Table E-13 summarize the results of the 840 system whole
blood and expired gas recovery precision testing.
Table EĆ11.āă840 Recovery and Precision TestingĊpH
Level
Mode
n
WRSD
*
Observed
Expected
Recovery
1
Syringe
15
0.002
7.251
7.251
100.0
Capillary
12
0.003
7.255
7.258
100.0
Microcapillary
17
0.006
7.261
7.256
100.1
Microsyringe
18
0.003
7.248
7.238
100.1
pH Only
15
0.004
7.235
7.235
100.0
2
Syringe
12
0.003
7.427
7.420
100.1
Capillary
14
0.003
7.429
7.423
100.1
Microcapillary
12
0.004
7.380
7.374
100.1
Microsyringe
15
0.002
7.400
7.397
100.0
pH Only
12
0.007
7.411
7.410
100.0
3
Syringe
15
0.002
7.663
7.654
100.1
Capillary
12
0.002
7.656
7.650
100.1
Microcapillary
18
0.007
7.664
7.664
100.0
Microsyringe
18
0.003
7.651
7.645
100.1
pH Only
15
0.004
7.636
7.635
100.0
* WRSD = within-run standard deviation