Defining reference and action ranges – Bayer HealthCare Rapidlab 800 User Manual

Page 427

background image

System Administration

5Ć17

Option 1

Option 2

tHb

ctHb

O

2

Cap

BO

2

AaDO

2

pO

2

(A–a)(T)

a/A

pO

2

(a/A)(T)

1.

Access the Parameter Names screen from the Menu screen:

a. Select

5 Operating

Setup

and press

Enter

.

b. Select

5 Parameter Names

and press

Enter

.

The Parameter Names screen appears, as shown in Figure 5-8.

Figure 5Ć8.ā Parameter Names Screen

Select the option you

want and press Enter.

2.

Select

Option 1

or

Option 2

and press Enter.

3.

Press

Done

when you finish.

4.

You can define another setup function or press

Exit Menu

to return to the

Ready screen.

Defining Reference and Action Ranges

Use this procedure to define the reference and action ranges for each parameter.
The system uses reference ranges to determine if a result is outside of the expected
range. The system uses action ranges to determine if a result requires immediate
action. Each laboratory should establish its own reference and action ranges for the
evaluation of patient results. Table 5-7 lists the default values for reference and
action ranges for the base models and for the base models with a CO-ox module.