Defining reference and action ranges – Bayer HealthCare Rapidlab 800 User Manual
Page 427

System Administration
5Ć17
Option 1
Option 2
tHb
ctHb
O
2
Cap
BO
2
AaDO
2
pO
2
(A–a)(T)
a/A
pO
2
(a/A)(T)
1.
Access the Parameter Names screen from the Menu screen:
a. Select
5 Operating
Setup
and press
Enter
.
b. Select
5 Parameter Names
and press
Enter
.
The Parameter Names screen appears, as shown in Figure 5-8.
Figure 5Ć8.ā Parameter Names Screen
Select the option you
want and press Enter.
2.
Select
Option 1
or
Option 2
and press Enter.
3.
Press
Done
when you finish.
4.
You can define another setup function or press
Exit Menu
to return to the
Ready screen.
Defining Reference and Action Ranges
Use this procedure to define the reference and action ranges for each parameter.
The system uses reference ranges to determine if a result is outside of the expected
range. The system uses action ranges to determine if a result requires immediate
action. Each laboratory should establish its own reference and action ranges for the
evaluation of patient results. Table 5-7 lists the default values for reference and
action ranges for the base models and for the base models with a CO-ox module.
5
Menu Code
5