Changing the auto clean time, Scheduling maintenance tasks – Bayer HealthCare Rapidlab 800 User Manual
Page 443

System Administration
5Ć33
Changing the Auto Clean Time
Use this procedure to change the time when the system performs the automatic
cleaning of the reagent manifold. Auto Clean occurs once every 24 hours. The
default time is 0200.
On the first day of the month, the system also prints out final statistical summary
reports of the previous month’s QC data at the time of the Auto Clean.
1.
Access the System Options screen from the Menu screen:
a. Press
6 System Setup
and press
Enter
.
b. Press
5 Systems Options
and press
Enter
.
The System Options screen appears, as shown in Figure 5-18.
Figure 5Ć18.ā System Options Screen for an 850
Move to the Auto Clean
Time field and enter the
time.
2.
Move to the Auto Clean Time field.
3.
Type the hour and minutes in the format, HH:MM. You can enter 0 – 23 hours
and 0 – 59 minutes.
4.
Press
Done
when you finish.
5.
Define another setup function or press
Exit Menu
to return to the Ready
screen.
Scheduling Maintenance Tasks
Use this procedure to:
S define the start date for scheduling maintenance on your system
S increase or decrease the frequency of the maintenance tasks
1.
Access the Maintenance Setup screen from the Menu screen:
a. Press
6 System Setup
and press
Enter
.