Maintaining the System
6.
Press
Continue
.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
7.
Press
No
.
8.
Calibrate the reagent pump from the Menu screen:
a. Select
3 Troubleshooting
and press
Enter
.
b. Select
1 Fluidics System
and press
Enter
.
c. Select
2 Pump Flow Rate
and press
Enter
.
d. Press
Calibrate Pump
.
e. Press
Exit Test
when the Pump calibration complete message appears.
9.
Perform a two-point calibration from the Menu screen:
a. Select
1 Calibration
and press
Enter
.
b. Select
2 Two-point
and press
Enter
.