Bayer HealthCare Rapidlab 800 User Manual
Page 186

3Ć18
Rapidlab 800 Operator's Manual
Figure 3Ć6.āăOpening the Measurement Module Door
Push up the door latches
and lift the door.
3.
Verify that the reference sensor fill solution is not below the fill line. Refer to
Filling the Reference Sensor
, page 3-69.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
4.
If necessary, fill the sensor as described in
Filling the Reference Sensor,
page 3-69.
5.
Verify that the pH, K
+
, Cl
–
, and Ca
++
sensors are almost full with only a
bubble at the top and that the Na
+
sensor is completely full.
If any sensor is not properly filled, replace the fill solution as described in
Filling the Measurement Sensors
, page 3-80.
6.
Close the measurement module door.
7.
Press
Continue
and allow the system to warm up.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
If . . .
Then . . .
you want to perform a
two-point calibration
press
Yes
.
you do not want to perform a
two-point calibration
a. Press
No
.
b. Press
Home
to return to the Ready screen.
!