4Ć71
Troubleshooting the System
Sensor
End Time, (sec)
Expected Value
Cal Signal
Expected Value
K
+
29.0 to 126.0 mV
15 to 90
Cl
–
29.0 to 126.0 mV
15 to 90
Ca
++
29.0 to 126.0 mV
20 to 90
Glu I
–5.35 to +16.45 nA
Glu A
–5.35 to +16.45 nA
45
Lac I
–5.35 to +16.45 nA
Lac A
–5.35 to +16.45 nA
45
Sensor
Slope Signal
Expected Value
End Time, (sec)
Expected Value
Sensor Slope Expected
Value
pH
217.1 to 443.3 mV
20 to 90
42.5 to 68.5 mV/Dec
pCO
2
–203.2 to +126.6 mV
25 to 90
42.5 to 68.5 mV/Dec
pO
2
–0.300 to +0.360 nA
5 to 90
0.002 to 0.020
nA/mmHg
Na
+
35.2 to 141.0 mV
15 to 90
42.5 to 68.5 mV/Dec
K
+
41.8 to 146.6 mV
15 to 90
42.5 to 68.5 mV/Dec
Ca
++
34.7 to 137.1 mV
20 to 90
19.0 to 36.8 mV/Dec
Cl
–
18.4 to 121.3 mV
15 to 90
–30.0 to –68.5 mV/Dec
Glu I
–5.35 to 16.45 nA
Glu A
1.85 to 88.45 nA
12 to 90
0.040 to 0.400
nA/mg/dL
Lac I
–5.35 to 16.45 nA
Lac A
–3.35 to 42.45 nA
12 to 90
1.00 to 13.00
nA/mmol/L
4.
Press
Exit Test
.