4Ć71

Troubleshooting the System

Sensor

End Time, (sec)

Expected Value

Cal Signal

Expected Value

K

+

29.0 to 126.0 mV

15 to 90

Cl

29.0 to 126.0 mV

15 to 90

Ca

++

29.0 to 126.0 mV

20 to 90

Glu I

–5.35 to +16.45 nA

Glu A

–5.35 to +16.45 nA

45

Lac I

–5.35 to +16.45 nA

Lac A

–5.35 to +16.45 nA

45

Sensor

Slope Signal

Expected Value

End Time, (sec)

Expected Value

Sensor Slope Expected

Value

pH

217.1 to 443.3 mV

20 to 90

42.5 to 68.5 mV/Dec

pCO

2

–203.2 to +126.6 mV

25 to 90

42.5 to 68.5 mV/Dec

pO

2

–0.300 to +0.360 nA

5 to 90

0.002 to 0.020
nA/mmHg

Na

+

35.2 to 141.0 mV

15 to 90

42.5 to 68.5 mV/Dec

K

+

41.8 to 146.6 mV

15 to 90

42.5 to 68.5 mV/Dec

Ca

++

34.7 to 137.1 mV

20 to 90

19.0 to 36.8 mV/Dec

Cl

18.4 to 121.3 mV

15 to 90

–30.0 to –68.5 mV/Dec

Glu I

–5.35 to 16.45 nA

Glu A

1.85 to 88.45 nA

12 to 90

0.040 to 0.400
nA/mg/dL

Lac I

–5.35 to 16.45 nA

Lac A

–3.35 to 42.45 nA

12 to 90

1.00 to 13.00
nA/mmol/L

4.

Press

Exit Test

.