Bayer HealthCare Rapidlab 800 User Manual
Page 549

Performance Characteristics
EĆ23
Level
Recovery
Expected
Observed
WRSD
*
n
Mode
4
Syringe
18
2.65
149.6
149.7
99.9
Capillary
22
0.83
150.0
149.7
100.2
Microcapillary
18
0.98
150.0
149.7
100.2
Microsyringe
18
0.62
149.4
149.7
99.8
Expired Gas
18
0.31
149.5
149.7
99.8
5
Syringe
21
1.40
377.4
377.9
99.9
Capillary
21
2.41
377.5
377.9
99.9
Microcapillary
18
5.50
378.9
377.9
100.3
Microsyringe
17
5.51
377.5
377.9
99.9
Expired Gas
17
1.55
379.2
377.9
100.3
* WRSD = within-run standard deviation