Bayer HealthCare Rapidlab 800 User Manual
Page 424

5Ć14
800 Series Operator's Manual
Parameter
System
Optional Units
Default Units
ctHb
g/dL
g/L, mmol/L
840, 850, 860
pH
pH
H
+
nmol/L
840, 850, 860
O
2
SAT, F
I
O
2
, sO
2
,
Hct, Hb Fractions
%
decimal fraction
840, 850, 860
Ca
++
mmol/L
mg/dL
850, 860
Glucose
mg/dL
mmol/L
860
Lactate
mmol/L
mg/dL
860
1.
Access the Units and Values screen from the Menu screen:
a. Select
5 Operating Setup
and press
Enter
.
b. Select
4 Units/Values
and press
Enter
.
The Units and Values screen appears, as shown in Figure 5-6.
Figure 5Ć6.ā Units and Values Screen for an 850
Press Next Screen
to view
additional parameters.
2.
Select the measurement units:
a. Move to the frame that contains the measurement unit you want to change.
b. Select the unit you want and press
Enter
.
c. Repeat steps a and b for each measurement unit you want to change.
d. Press
Next Screen
to view additional parameters.
e. Repeat steps a and b for each measurement unit you want to change.
3.
Press
Done
.
4.
You can define another setup function or press
Exit Menu
to return to the
Ready screen.
5
Menu Code
4