Bayer HealthCare Rapidlab 800 User Manual
Page 577

Performance Characteristics
EĆ51
Table EĆ46.āă800 COĆox Module Recovery and Precision with Total Hemoglobin of
18-22 g/dL and Oxyhemoglobin > 80%
Analyte
Mode
n
Mean
Reference
Mean Difference
from Reference
WRSD
*
tHb
Capillary
72
21.04
0.21
0.21
Syringe
68
20.72
0.19
0.22
FO
2
Hb
Capillary
72
90.03
–0.73
0.24
Syringe
68
92.31
–1.21
0.64
FCOHb
Capillary
72
6.56
0.53
0.16
Syringe
68
4.48
0.95
0.72
FMetHb
Capillary
72
3.41
–0.08
0.21
Syringe
68
3.21
–0.01
0.20
FHHb
Capillary
72
0.00
0.27
0.11
Syringe
68
0.00
0.28
0.16
* WRSD = within-run standard deviation