Bayer HealthCare Rapidlab 800 User Manual
Page 544

EĆ18
800 Series Operator's Manual
Parameter
TotSD
[
WRSD
*
Mean
n
Level
2
71
133.0
0.21
0.86
3
72
154.5
0.36
1.58
K
+
1
72
2.42
0.007
0.023
2
71
4.96
0.017
0.036
3
72
7.71
0.022
0.087
Cl
–
1
72
79.2
0.35
0.57
2
71
101.7
0.27
0.53
3
72
121.6
0.29
0.92
Ca
++
1
72
1.54
0.006
0.012
2
71
1.12
0.005
0.011
3
72
0.62
0.004
0.009
* WRSD = within-run standard deviation
[
TotSD = total standard deviation
Precision on aqueous calibration verification materials was estimated using four
850 systems. As many as seven runs per instrument were made over seven days.
Two replicates of each control level were analyzed in each run.
Table E-15 summarizes the results of the 850 system precision for CVM levels 1
and 4.
Table EĆ15.āă850 CVM Precision Results
Parameter
Level
n
Mean
WRSD
*
TotSD
[
pH
1
44
6.806
0.001
0.004
4
44
7.805
0.002
0.004
pCO
2
1
44
103.5
0.93
1.49
4
44
13.4
0.13
0.28
(Continued)