D19 fluid detector error – Bayer HealthCare Rapidlab 800 User Manual
Page 316
4Ć32
800 Series Operator's Manual
D19 Fluid Detector Error:
Problem
Fluid detector millivolt (mV) reading is beyond predefined limits.
Qualifier:
FD5
Solutions
Check the CO-ox sample path for obstructions. If you see any obstructions,
remove them as described in Removing Obstructions from the CO-ox Sample
Path, page 4-87.
Check the sample tubing that goes through FD5 for cracks, leaks, and
discoloration. If you see cracks or leaks, replace the tubing as described in
Replacing the CO-ox Sample Tubing in Section 3.
Perform the Fluid Detector Test as described in Fluid Detector Test, page 4-66.
Deproteinize the sample path as described in Deproteinizing the Sample Path in
Section 3 and repeat the Fluid Detector Test.
If the D code reappears, contact your Service Representative.
D19