Replacing the reference sensor – Bayer HealthCare Rapidlab 800 User Manual
Page 243

3Ć71
Maintaining the System
Replacing the Reference Sensor
Materials required:
S reference sensor replacement kit
S hex tool
S lint-free tissue
BIOHAZARD:
Refer to Appendix A, Protecting Yourself from Biohazards, for
recommended precautions when working with biohazardous materials.
1.
Stop the system from the Menu screen:
a. Select
2
Maintenance
and press
Enter
.
b. Select
7 Stop System
and press
Enter
.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
2.
Remove the reference sensor as shown in Figure 3-43:
a. Push up the latches on the measurement module door and lift the door.
b. Push the spring-loaded latch to the right.
c. Grasp the tab on the reference sensor and pull the sensor up and out of the
measurement module.
d. Discard the reference sensor according to your laboratory protocol.
Figure 3Ć43.āăRemoving the Reference Sensor from an 850
Grasp the tab on the reference
sensor and pull it up and out of the
measurement module.
Push the springĆloaded latch
to the right.
3.
Fill the electrode compartment in the new reference sensor:
a. Remove the reference sensor cassette (KCl reservoir) from its box.
b. Insert the tip of the KCl fill solution container into the electrode
compartment as shown in Figure 3-44.