800 Series Operator's Manual
5Ć56
Printing a QC Setup Report
Use this procedure to print a QC setup report for any QC file.
1.
Access the QC File screen from the Menu screen:
a. Select
5 Operating Setup
and press
Enter
.
b. Select
1 QC Setup
and press
Enter
.
c. Select
1 QC Files
and press
Enter
.
2.
Select the file number that you want to print.
3.
Press
Print QC Setup
.
The QC Setup report prints. Figure 5-34 shows an example of a QC Setup
Report.
Figure 5Ć34.ā QC Setup Report
............. . . . . . . .
. . . . . . . . . . . . . . . . . . . .
.........................
840
850
QC SETUP REPORT APR 12 1994
11:42
SYSTEM 850-1001
GLOBAL QC SETUP
Auto ID is: On
Auto accept QC: Off
QC FILE SETUP
File 1
QC ID: 473842
Level: 2
Lot number: 41561
Expiration date: Dec 30 95
MEAN TARGET RANGE ACTION RANGE
pH 7.419 ( 7.399 - 7.439)( 7.389 - 7.449)
pCO
2
43.4 ( 38.4 - 48.4)( 35.9 - 50.9)
pO
2
103.1 ( 96.1 - 110.1)( 92.6 - 113.6)
Na+ 135.0 ( 130.0 - 140.0)( 127.5 - 142.5)
K+ 4.81 ( 4.31 - 5.31)( 4.06 - 5.56)
Ca++ 1.16 ( 1.06 - 1.26)( 1.01 - 1.31)
Cl- 102 ( 97 - 107)( 94.5 - 109.5)
Glucose 102.5 ( 97.5 - 107.5)( 87.5 - 117.5)
Lactate 0.92 ( 0.71 - 1.11)( 0.61 - 1.21)
ctHb 14.5 ( 13.5 - 15.5)( 13.0 - 16.0)
FO
2
Hb 1.2 ( 0.7 - 1.7)( 0.4 - 2.0)
FCOHb 95.5 ( 93.5 - 97.5)( 92.5 - 98.5)
FMetHb 0.4 ( 0.2 - 0.6)( 0.1 - 0.7)
FHHb 2.6 ( 2.1 - 3.1)( 1.8 - 3.4)
............................... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
860
These results
appear when an
845, 855, or 865
is interfaced.
..............
.............. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CO-ox
4.
Press
Menu
.
5.
You can define another setup function or press
Exit Menu
to return to the
Ready screen.
1
1
Menu Code