CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
2.
Remove the sensor as shown in Figure 3-59:
a. Push up the latches on the measurement module door and lift the door.
Figure 3Ć59.āăRemoving a Measurement Sensor from an 850
Grasp the tab on the sensor and pull the
sensor up and out of the measurement
module.
Push the springĆloaded latch
to the right.
b. Push the spring-loaded latch to the right.
c. Grasp the tab on the sensor and pull the sensor up and out of the
measurement module.
d. Discard the sensor.
3.
Perform the appropriate action.
If you are replacing . . .
Then . . .
the chloride sensor
go to step 4.
other measurement sensors or the
sample ground/temperature sensor
go to step 5.
4.
Fill the chloride sensor with the Na
+
, K
+
, Cl
–
, Ca
++
fill solution:
a. Remove the new sensor from the box.
CAUTION:ĄĂ
Do not touch the internal electrode wire. The wire is fragile and is
easily damaged.
b. Unscrew the internal electrode and set it aside on a lint-free tissue.
c. Rinse the electrode compartment with 3 drops of fill solution and then empty
it.
d. Fill the sensor almost full, leaving a bubble at the top.
!
860