Page 273

3Ć101
Maintaining the System
a. Select
2 Maintenance
and press
Enter
.
b. Select
7 Stop System
and press
Enter
.
2.
Ensure that the sample probe is fully retracted.
3.
Remove the old capillary seal:
a. Grasp the top of the capillary seal and pull it toward the right and out of the
sample port, as shown in Figure 3-69.
Figure 3Ć69.āăReplacing the Capillary Seal
Capillary Seal
Sample Port
b. Discard the capillary seal according to your laboratory protocol.
4.
Install the new capillary seal:
a. Hold the seal at the top, angle it toward the right, and push it into the
opening in the sample port.
b. Insert a capillary tube into the port to ensure that the capillary seal fits
securely in place.
5.
Press
Continue
.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
6.
Press
Yes
to perform a two-point calibration.