2Ć23

Operating the System

If the patient ID entered at the 800 system and the 270 CO-oximeter do not match,
the Patient ID Mismatch message box appears.

If . . .

Then . . .

you want to combine
the 270 results with
the 800 system results

a. Press

Combine Results

. The Patient Information screen

appears.

b. Type the patient ID and press

Done

.

you do not want to
combine the 270
results with the 800
system results

press

Cancel

.

Separate reports are generated at the 270 COĆoximeter

and at the 800 system.

If the 800 system completes analysis before the 270 CO-oximeter completes
analysis, the Waiting for CO-oximetry Results message box appears.

If . . .

Then . . .

you want to combine
results

press

Combine Results

. The 800 system creates a combined

report when the results are ready.

you do not want to
combine results

press

Cancel

. Separate reports will be generated at the 270

CO-oximeter and at the 800 system.

Procedural

Notes