Bayer HealthCare Rapidlab 800 User Manual
Page 121

2Ć23
Operating the System
If the patient ID entered at the 800 system and the 270 CO-oximeter do not match,
the Patient ID Mismatch message box appears.
If . . .
Then . . .
you want to combine
the 270 results with
the 800 system results
a. Press
Combine Results
. The Patient Information screen
appears.
b. Type the patient ID and press
Done
.
you do not want to
combine the 270
results with the 800
system results
press
Cancel
.
Separate reports are generated at the 270 COĆoximeter
and at the 800 system.
If the 800 system completes analysis before the 270 CO-oximeter completes
analysis, the Waiting for CO-oximetry Results message box appears.
If . . .
Then . . .
you want to combine
results
press
Combine Results
. The 800 system creates a combined
report when the results are ready.
you do not want to
combine results
press
Cancel
. Separate reports will be generated at the 270
CO-oximeter and at the 800 system.
Procedural
Notes