3Ć105

Maintaining the System

b. Select

3 Measurement

and press

Enter

.

c. Select

2 COox Optics

and press

Enter

.

The COox Optics Test screen appears.

13. Press

Start Test

.

14. Check the screen for the message, Lamp test passed.

15. Press

Exit Test

.

16. Press

Exit Menu

.

17. Perform a one-point calibration from the Menu screen:

a. Select

1 Calibration

and press

Enter

.

b. Select

1 One-point

and press

Enter

.

Replacing the System Fuses

Materials required:

S 2 fuses of the appropriate rating
S small, flat-blade screwdriver

Refer to Table 3-2 to identify the correct fuses for the voltage you use.

Table 3Ć2.āăSystem Fuse Selection

Voltage

Fuse Rating

Fuse Type

100/120V

4A Slo Blo

5 x 20 mm

220/240V

2A Slo Blo

5 x 20 mm

WARNING

To prevent electrical shock or damage to the system, remove power

from your system before performing this procedure.

1.

Shut down the system from the Menu screen:

a. Select

7 System Utilities

and press

Enter

.

b. Select

3 Shutdown

and press

Enter

.

c. Press

Yes

.

CAUTION:

You must wait at least 1 minute before you disconnect the power cord,

and then wait at least 10 seconds before you reconnect the power cord. If you do
not adhere to the time intervals, you can damage the system.

2.

Wait until prompted and then disconnect the power cord from the power
source.

1

Menu Code

1

(from the Main Menu)

7

Menu Code

3

!