Measurement test, Printing stored signals – Bayer HealthCare Rapidlab 800 User Manual
Page 356
4Ć70
800 Series Operator's Manual
4.
Perform the appropriate action.
If you see the . . .
Then . . .
Temperature control system on
message
a. Press
Stop Test
.
b. Look at the status area to see if the D38
code has disappeared.
c. Allow the system to reach operating
temperature.
Temperature control system off
message
the system cannot reset the temperature control
system.
5.
Press
Exit Test
.
Measurement Test
Use this procedure to test the measurement electronics. You can:
S print stored signals from the last successful two-point calibration
S test the sensor circuit without a sensor and with a test/blank sensor
Printing Stored Signals
1.
Access the Measurement Test from the Menu screen:
a. Select
3 Troubleshooting
and press
Enter
.
b. Select
3 Measurement
and press
Enter
.
c. Select
1 Measurement
and press
Enter
.
2.
Press
Print Last Cal Mv
.
3.
Compare the results on the report to the expected values in the following
tables.
Sensor
Cal Signal
Expected Value
End Time, (sec)
Expected Value
pH
194.0 to 406.0 mV
20 to 90
pCO
2
–216.0 to +106.0 mV
25 to 90
pO
2
–0.129 to +2.071 nA
5 to 90
Na
+
29.0 to 131.0 mV
15 to 90
(Continued)