Defining parameter values – Bayer HealthCare Rapidlab 800 User Manual
Page 425

System Administration
5Ć15
If you change parameter units for pH after the system has started to collect data,
QC statistics are still computed in the current units.
Defining Parameter Values
Use this procedure to define the values for ctHb and p50 that the system uses when
no value is entered or measured, and to define the value for the O
2
binding factor
and ctO
2
(a-v), which are constants supplied by the system. The system uses these
values for certain calculations. Table 5-5 lists the default values and valid ranges
for ctHb, O
2
binding factor, p50, and ctO
2
(a-v).
Table 5Ć5.āăDefault Measurement Values
Parameter
Valid Range
Default Value
Units
ctHb
2.0 – 25.0
20 – 250
1.2 – 15.5
15.0
g/dL
g/L
mmol/L
p50
0.0 – 100.0
0.00 – 13.32
26.5
mmHg
kPa
O
2
Binding
1.35 – 1.39
1.39
ctO
2
(a-v)
0.0 – 20.0
0 – 200
0.0 – 9.0
3.5
mL/dL
mL/L
mmol/L
1.
Access the Units and Values screen from the Menu screen:
a. Select
5 Operating Setup
and press
Enter
.
b. Select
4 Units/Values
and press
Enter
.
The first Units and Values screen appears.
2.
Press
Next Screen
.
The second Units and Values screen appears as shown in Figure 5-7.
Procedural