Interrupting calibrations – Bayer HealthCare Rapidlab 800 User Manual
Page 158

2Ć60
Rapidlab 800 Operator's Manual
Interrupting Calibrations
Press
Cancel
to interrupt a calibration.
If you interrupt an automatic calibration, the system will attempt to start the
calibration again in 90 seconds. You can delay calibrations for up to 30 minutes
beyond the scheduled time. After 30 minutes, the system starts the required
calibration. You cannot analyze samples until the required calibration completes.
You cannot defer a one-point calibration that is initiated following rapid sampling.
You can interrupt or defer the one-point calibration that occurs when the 5 minute
timer expires.
The maximum times between automatic one-point and two-point calibrations are
60 minutes and 240 minutes, respectively. Five minutes before an automatic
calibration is scheduled, a status message appears on the screen indicating that a
calibration is pending.
Procedural
Notes
860