Bayer HealthCare Rapidlab 800 User Manual
Page 458

800 Series Operator's Manual
5Ć48
9.
Press
Done
.
A message appears prompting you to set up another QC file.
If you want to . . .
Then press . . .
save your entries and
create another QC file
Yes
. The QC File screen appears. Continue with step 2.
save your entries and you
do not want to create
another file
No
.
The Menu screen appears. Define another setup
function or press
Exit Menu
to return to the Ready
screen.
return to the QC File
Setup (ranges) screen
Cancel
.
The cursor returns to the last field you
completed.
If you press Done before you complete both the low and high fields for a
parameter, a message appears telling you the empty field is an invalid range.
Press
OK
. Complete the field. If any other incomplete fields exist, the cursor
moves to those fields.
If you press Menu before you press Done, a message appears prompting you
to save your entries.
If you want to . . .
Then press . . .
save your entries
Yes
. The Menu screen appears.
delete your entries
No
. The Menu screen appears.
return to the screen
Cancel
.
The cursor returns to the last field you
completed.
Procedural
Notes