4Ć7

Troubleshooting the System

System Message

Description

COox Sample
Chamber Temp
Error

Appears when the CO-ox sample chamber temperature is not in
range. The system cannot accept tHb sample measurement
requests. See Troubleshooting System Messages, page 4-114.

COox Sample
Temp Out of Range

Appears when the CO-ox sample chamber temperature is not in
range at the end of measurement sequence. See Troubleshooting
System Messages,
page 4-114.

Correlation
Adjustment

Appears when the correlation coefficient is changed.

Data Entry
Incomplete

Appears when required data entry fields were not completed.

Device Connected
to Port __

Appears after you configure an 800 system for an external device
and the connection is established.

Excessive Bubbles
in COox Sample

Appears when the system detects a non-continuous fluid (sample
or QC material) in the sample chamber and cannot complete the
analysis. See Troubleshooting System Messages, page 4-115.

Excessive Bubbles
in COox Zero

Appears when the system detects a non-continuous fluid (7.3
buffer) in the sample chamber and cannot perform the zero. See
Troubleshooting System Messages, page 4-115.

NOTE:

The CO-ox module is zeroed during the loading of the

7.3 buffer during a one- or two-point calibration.

Excessive Bubbles
in tHb Slope

Appears when the system detects a non-continuous fluid (slope
material) in the sample chamber and cannot slope the CO-ox
module. See Troubleshooting System Messages, page 4-115.

Excessive Scatter in
COox Meas

Appears when the lipid level detected in the CO-ox sample
exceeds the expected physiological maximum. Also appears if
the hemolyzer is disconnected. See Troubleshooting System
Messages,
page 4-115.

File (file number)
Outdated QC

Notifies you that a specific QC file is outdated.

If Blood,
Question Data

Optical measurements indicate that the CO-oximeter results
should be reviewed. A question mark (?) is printed next to the
CO-oximeter results on reports. See Troubleshooting System
Messages,
page 4-116.

(Continued)