Bayer HealthCare Rapidlab 800 User Manual
Page 114

2Ć16
Rapidlab 800 Operator's Manual
Figure 2Ć10.ā Sample Type Menu
Select pH Only.
3.
Select
pH Only
and press
Enter
.
4.
Press
Analyze
.
NOTE:ĄĂ
Ensure that the trailing edge of the sample remains in contact with the
sample ground/temperature sensor.
5.
Turn the sample pump clockwise until the leading edge of the sample fills the
reference sensor, as shown in Figure 2-11.
Figure 2Ć11.ā Sample at Reference Sensor on an 850
Reference Sensor
Sample Ground/
Temperature Sensor
Sample Pump
6.
Press
Sample in Place
.