Bayer HealthCare Rapidlab 800 User Manual
Page 564

EĆ38
800 Series Operator's Manual
Level
Recovery
Expected
Observed
WRSD
*
n
Mode
Capillary
23
0.44
50.3
49.9
100.8
Microcapillary
24
0.71
49.9
49.9
100.0
Microsyringe
24
0.54
49.8
49.9
99.7
Expired Gas
18
0.62
49.4
49.9
99.1
5
Syringe
24
0.52
71.3
71.3
100.1
Capillary
24
0.82
71.2
71.3
99.8
Microcapillary
24
1.07
71.0
71.3
99.5
Microsyringe
24
1.46
71.0
71.3
99.5
Expired Gas
24
0.24
70.5
71.3
98.8
* WRSD = within-run standard deviation
Table EĆ33.āă860 Recovery and Precision TestingĊpO
2
Level
Mode
n
WRSD
*
Observed
Expected
Recovery
1
Syringe
24
0.21
28.5
28.5
99.9
Capillary
24
0.25
28.5
28.5
100.1
Microcapillary
24
0.44
28.6
28.5
100.4
Microsyringe
24
0.33
28.7
28.5
100.7
Expired Gas
24
0.12
28.0
28.5
98.1
2
Syringe
24
0.35
50.2
49.9
100.6
Capillary
23
0.23
49.9
49.9
100.1
Microcapillary
24
0.38
50.0
49.9
100.2
Microsyringe
24
0.49
49.7
49.9
99.6
Expired Gas
24
0.56
49.8
49.9
99.7
(Continued)