Cleaning the sample ground/temperature sensor – Bayer HealthCare Rapidlab 800 User Manual
Page 219

3Ć49
Maintaining the System
Cleaning the Sample Ground/Temperature Sensor
Materials required:
S 0.022 inch diameter clot-removal line (to fit through a 0.030 inch ID pathway)
1.
Stop the system from the Menu screen:
a. Select
2
Maintenance
and press
Enter
.
b. Select
7 Stop System
and press
Enter
.
2.
Push up the latches on the measurement module door and lift the door.
3.
Push the spring-loaded latch to the right as shown in Figure 3-31.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
4.
Grasp the tab on the sample ground/temperature sensor and pull the sensor up
and out of the measurement module.
Figure 3Ć31.āăRemoving the Sample Ground/Temperature Sensor from an 850
Push the springĆloaded
latch to the right.
Grasp the tab on the sample ground
sensor and pull it up and out of the
measurement module.
5.
Push a clot removal line through the sensor to clear away any obstructions as
shown in Figure 3-32.
Figure 3Ć32.āăCleaning the Sample Ground/Temperature Sensor
Clot Removal Line
OĆring
2
Menu Code
7
!