D35 electronics error, Problem, Solutions – Bayer HealthCare Rapidlab 800 User Manual
Page 323

4Ć39
Troubleshooting the System
D35 Electronics Error:
Problem
The system detects an error in the electronics system.
Qualifiers:
1 2
4
5
6
7
8
9
10 11 12 13
Solutions
Qualifier
Procedure
1, 2, 4, 5, 8,
9, 11
perform the Measurement Test as described on page 4-70.
If the measurement test does not remove the D35, shut the system
down and restart it as described in Shutting Down and Restarting the
System in Section 5.
6 or 7
shut down and restart the system as described in Shutting Down and
Restarting the System in Section 5.
10
check the temperature control as described in Temperature/pAtm Test,
page 4-69.
12
perform the Valves Test as described on page 4-62.
13
perform the Pump Functions Test as described on page 4-58.
If the D code reappears, contact your Service Representative.
D35