3.
Deproteinize the sample path:
a. Prepare the deproteinizer as directed on the package.
b. Select
2 Maintenance
and press
Enter
.
c. Select
1 Deproteinize
and press
Enter
.
d. Invert the deproteinizer vial several times to mix.
e. Insert an aspiration adapter into the sample port and insert the other end
into the deproteinizer.
f. Press
Analyze
.
g. When prompted, remove the adapter.
h. Wait for the deproteinizing cycle to finish.
A message box appears prompting you to condition the sensors.
4.
Perform the cleaning cycle:
a. Press
Yes
.
b. Insert an aspiration adapter into the sample port and immerse the other end
in the glutaraldehyde solution.
c. Press
Analyze
.
d. When prompted, remove the adapter.
e. Wait 5 minutes for the cleaning cycle to finish.
NOTE:ĄĂ
Press
Cancel
if you want to stop cleaning.
The cleaning cycle finishes. The system performs an extended wash sequence.
The Calibrate System message box appears at the end of the wash sequence.
5.
Press
No
.
6.
Initiate a wash sequence from the Menu screen:
a. Select
2 Maintenance
and press
Enter
.
b. Select
8 Wash
and press
Enter
.
NOTE:ĄĂ
As you install the biosensors, ensure that the biosensors are in the
correct location. Visually verify that you align the contacts on the biosensors
with the contacts in the measurement module.
CAUTION:ĄĂ
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
7.
When the wash sequence finishes, remove the test/blank glucose and lactate
sensors and reinstall the biosensors as described in
Replacing the Glucose and
Lactate Biosensors
page 3-86.
2
Menu Code
1
2
Menu Code
8
(from the Main Menu)
!