Replacing the Reagent Manifold Vent Filter

Perform the following procedure annually. This schedule is based on analyzing 30
samples per day. If your laboratory analyzes more than 30 samples per day,
perform this maintenance more frequently.

Materials required:

S reagent manifold vent filter
1.

Remove the old reagent manifold vent filter from the reagent manifold vent
inlet and discard it.

2.

Insert the new filter into the reagent manifold vent inlet as shown in
Figure 3-26.

Figure 3Ć26ā.ăReplacing the Reagent Manifold Vent Filter

Reagent Manifold
Vent Filter

3.

Press

Home

to return to the Ready screen.