Bayer HealthCare Rapidlab 800 User Manual
Page 230

3Ć60
Rapidlab 800 Operator's Manual
NOTE:
If your system has a CO-ox module, follow the procedure described for
the appropriate base model. For example, information identified for an 860 also
applies to an 865.
CAUTION:
Do not remove or return the sensors to the measurement module
without first discharging static buildup. Touch the inner surface of the module
frame to discharge static buildup.
CAUTION:
Do not expose the glucose and lactate biosensors to deproteinizer or
bleach. Replace the biosensors with test/blank sensors (TB4) before cleaning the
sample path. Reinstall the biosensors within 2 hours.
1.
Take the appropriate action.
If you have an . . .
Then . . .
840 or 850
go to step 2.
860
a. Remove the glucose and lactate biosensors.
b. Install the test/blank glucose and lactate sensors
(TB4).
c. Go to step 2.
CAUTION:ĄĂ
Do not expose the reference sensor to bleach. Replace the
reference sensor with the test/blank reference sensor (TB5). Do not substitute a
new reference sensor.
2.
Replace the reference sensor with a test/blank reference sensor:
a. Remove the reference sensor from the measurement module and set it
aside.
b. Install a test/blank reference sensor (TB5) in the measurement module.
3.
Deproteinize the sample path:
a. Prepare the deproteinizer as directed on the package.
b. Select
2 Maintenance
and press
Enter
.
c. Select
1 Deproteinize
and press
Enter
.
d. Invert the deproteinizer vial several times to mix.
e. Insert an aspiration adapter into the sample port and insert the other end
into the deproteinizer.
f. Press
Analyze
.
g. Remove the adapter when prompted.
h. Wait 5 minutes for the deproteinizing cycle to finish.
A message box appears prompting you to condition the sensors.
!
!
860
!
2
Menu Code
1