Selecting parameter panels – Bayer HealthCare Rapidlab 800 User Manual
Page 127

2Ć29
Operating the System
Selecting Parameter Panels
Use this procedure when you want to measure specific parameters for a single
patient sample. After the sample is analyzed, the system restores the default panel
for your system. Different panels are available for each system.
1.
Press
Change Parameters
.
The Parameter Panels menu appears, as shown in Figure 2-16.
Figure 2Ć16.ā Parameter Panels Menu for an 850 System
Select the parameter
panel you want.
2.
Select a parameter.
If you want to select . . .
Then . . .
one of the panels displayed
select the parameter panel and press
Enter
.
Custom
panels are listed as Panels 1–5 and are defined during
system setup by the system administrator or authorized
personnel.
specific parameters to
analyze
a. Press
Select Parameters
.
The Select Parameters
screen appears.
b. Select the parameters you want and press
Done
to
return to the Ready screen.
3.
Insert the sample device into the sample port.
4.
Press
Analyze
.
After the next sample is analyzed, the default panel is restored.
If you select Previous Screen, the system returns to the Ready screen and restores
the default panel.
Procedural
Notes