Bayer HealthCare Rapidlab 800 User Manual
Page 542

EĆ16
800 Series Operator's Manual
Level
Recovery
Expected
Observed
WRSD
*
n
Mode
4
Syringe
24
0.97
150.5
149.7
100.5
Capillary
18
0.86
150.2
149.7
100.3
Microcapillary
24
2.00
151.8
149.7
101.4
Microsyringe
24
0.80
150.6
149.7
100.6
Expired Gas
24
0.25
150.4
149.7
100.5
5
Syringe
24
1.20
377.9
377.9
100.0
Capillary
18
2.40
379.0
377.9
100.3
Microcapillary
24
4.08
380.7
377.9
100.7
Microsyringe
24
3.22
378.9
377.9
100.3
Expired Gas
24
0.66
378.7
377.9
100.2
* WRSD = within-run standard deviation