beautypg.com

Bayer HealthCare Rapidlab 800 User Manual

Page 576

background image

EĆ50

800 Series Operator's Manual

Analyte

WRSD

*

Mean Difference

from Reference

Mean

Reference

n

Mode

Syringe

66

92.18

0.90

0.39

FCOHb

Capillary

72

4.23

–0.43

0.33

Syringe

66

4.24

–0.45

0.12

FMetHb

Capillary

72

3.81

–0.95

0.68

Syringe

66

3.58

–0.67

0.39

FHHb

Capillary

72

0.00

0.23

0.08

Syringe

66

0.00

0.22

0.05

* WRSD = within-run standard deviation

Table EĆ45.āă800 COĆox Module Recovery and Precision with Total Hemoglobin of

10-16 g/dL and Oxyhemoglobin > 80%

Analyte

Mode

n

Mean

Reference

Mean Difference

from Reference

WRSD

*

tHb

Capillary

68

15.18

–0.36

0.35

Syringe

66

15.00

–0.31

0.11

FO

2

Hb

Capillary

68

92.31

–0.09

0.25

Syringe

66

92.27

–0.23

0.29

FCOHb

Capillary

68

4.52

0.24

0.45

Syringe

66

4.02

0.19

0.19

FMetHb

Capillary

68

3.17

–0.44

0.25

Syringe

66

3.71

–0.76

0.20

FHHb

Capillary

68

0.00

0.29

0.08

Syringe

66

0.00

0.34

0.08

* WRSD = within-run standard deviation