Bayer HealthCare Rapidlab 800 User Manual
Page 576
EĆ50
800 Series Operator's Manual
Analyte
WRSD
*
Mean Difference
from Reference
Mean
Reference
n
Mode
Syringe
66
92.18
0.90
0.39
FCOHb
Capillary
72
4.23
–0.43
0.33
Syringe
66
4.24
–0.45
0.12
FMetHb
Capillary
72
3.81
–0.95
0.68
Syringe
66
3.58
–0.67
0.39
FHHb
Capillary
72
0.00
0.23
0.08
Syringe
66
0.00
0.22
0.05
* WRSD = within-run standard deviation
Table EĆ45.āă800 COĆox Module Recovery and Precision with Total Hemoglobin of
10-16 g/dL and Oxyhemoglobin > 80%
Analyte
Mode
n
Mean
Reference
Mean Difference
from Reference
WRSD
*
tHb
Capillary
68
15.18
–0.36
0.35
Syringe
66
15.00
–0.31
0.11
FO
2
Hb
Capillary
68
92.31
–0.09
0.25
Syringe
66
92.27
–0.23
0.29
FCOHb
Capillary
68
4.52
0.24
0.45
Syringe
66
4.02
0.19
0.19
FMetHb
Capillary
68
3.17
–0.44
0.25
Syringe
66
3.71
–0.76
0.20
FHHb
Capillary
68
0.00
0.29
0.08
Syringe
66
0.00
0.34
0.08
* WRSD = within-run standard deviation