Fluid detector test – Bayer HealthCare Rapidlab 800 User Manual
Page 352

4Ć66
800 Series Operator's Manual
Fluid Detector Test
Use this procedure to test the ability of the fluid detectors to detect the presence of
fluids and to detect whether the fluids are clear or opaque. You can also test fluid
detectors 1, 1A, and 2 with whole blood by inserting a sample device containing
whole blood into the sample port.
1.
Access the Fluid Detector Test from the Menu screen:
a. Select
3 Troubleshooting
and press
Enter
.
b. Select
1 Fluidics System
and press
Enter
.
c. Select
5 Fluid Detector
and press
Enter
.
The Fluid Detector Test screen appears as shown in Figure 4-8.
Figure 4Ć8.ā Fluid Detector Test Screen
Perform the
Baseline/Clear test.
Check here for results.
2.
Test the fluid detectors with clear fluid and air:
a. Select
Baseline/Clear
and press
Enter
.
b. Press
Start Test
.
c. Check the screen for the message FD1, FD1A, FD2, FD3, FD4, FD5
acceptable to verify correct fluid detector function.
d. Press
Exit Test
.
3.
Perform the appropriate action.
If . . .
Then . . .
the fluid detector failed the Baseline test
continue with step 6.
you want to test whole blood
continue with step 4.
you want to return to the Menu screen
continue with step 6.
5
1
Menu Code
3