Semiannual Maintenance
Perform the following procedures every 6 months. The semiannual maintenance
described in this section is based on analyzing 30 samples per day. If your
laboratory analyzes more than 30 samples per day, perform this maintenance more
frequently.
Replacing the Air Filter
Materials required:
S air filter
1.
Push the air filter cover down until it snaps out and then pull the cover away
from the system as shown in Figure 3-17.
Figure 3Ć17ā.ăReplacing the Air Filter
Air Filter
Air Filter Cover
2.
Remove the old air filter and discard it.
3.
Press the new air filter into place.