3Ć95

Maintaining the System

4.

Disconnect the tubing from the gas regulator and discard the tubing.

5.

Connect one end of the new tubing to the fitting on the regulator.

6.

Connect the other end of the tubing to the fitting on the reagent manifold.

NOTE:

The average main tank pressure is 2200 psi. The average secondary valve

pressure is 3 to 5 psi.

7.

Slowly open the gas tank by turning the valve stem counterclockwise with the
wrench.

Turn the valve stem approximately 3/4 turn until the regulator pressure gauge
indicator stops rising.

8.

Open the valve stem one more turn.

9.

Listen carefully for any gas leaks.

10. Check for leaks using soapy water and watching for bubbles.

11. Verify that the regulator outlet gauge indicates 3 to 5 psi.

12. Press

Continue

.

A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.

13. Press

No

.

14. Check the gas flow rate from the Menu screen:

a. Select

3 Troubleshooting

and press

Enter

.

b. Select

1 Fluidics System

and press

Enter

.

c. Select

4 Valves

and press

Enter

.

d. Select the gas type that you replaced and press

Enter

.

e. Insert an aspiration adapter into the sample port and immerse the open end

of the adapter into a small container of reagent water.

f. Press

Start Test

.

g. Verify that a steady stream of bubbles flows into the water.

h. Press

Stop Test

.

i. Verify that the bubbles stop flowing into the water.

j. Remove the adapter.

k. Press

Exit Test

.

15. Perform a two-point calibration from the Menu screen:

a. Select

1 Calibration

and press

Enter

.

b. Select

4 Gas Two-point

and press

Enter

.

4

1

Menu Code

3

(from the Main Menu)

1

Menu Code

4

(from the Main Menu)