Rapidlab 800 Operator's Manual
Figure 3Ć52.āăRemoving the Internal Electrode
Internal Electrode
OĆring
Electrode
Compartment
OĆring
4.
Discard the internal electrode according to your laboratory protocol.
5.
Empty the KCl fill solution from the cassette.
6.
Add 3 drops of KCl fill solution to the electrode compartment and then drain
the compartment.
7.
Fill the electrode compartment in the new cassette as shown in Figure 3-53:
a. Insert the tip of the KCl fill solution container into the electrode
compartment in the cassette.
b. Partially fill the electrode compartment by gently squeezing the KCl fill
solution container.
c. Tap the front face of the sensor with your knuckle to release any bubbles.
d. Continue filling until the KCl fill solution enters the KCl reservoir.