Bayer HealthCare Rapidlab 800 User Manual
Page 279

3Ć107
Maintaining the System
12. Reconnect the power cord to the power source and allow the system to warm
up for at least 15 minutes.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
13. Press
No
.
14. Verify the system component temperatures from the Menu screen:
a. Select
3 Troubleshooting
and press
Enter
.
b. Select
2 Temp/pAtm
and press
Enter
.
c. Press
Start Test
.
d. Verify that the system components warm up.
e. Press
Stop Test
.
f. If the temperature control system is off, press
Reset Control
and allow the
system to warm up.
g. Press
Exit Test
.
15. Perform a two-point calibration from the Menu screen:
a. Select
1 Calibration
, and press
Enter
.
b. Select
2 Two-point
, and press
Enter
.
3
Menu Code
2
1
Menu Code
2
(from the Main Menu)