Bayer HealthCare Rapidlab 800 User Manual
Page 269
3Ć97
Maintaining the System
Figure 3Ć65.āăRemoving the Reagent Pump Roller Cage
Platen
Roller Cage
Roller Cage Shaft
Large Tubing
Small Tubing
Tubing Cuff
6.
Clean the roller cage shaft with a lint-free tissue moistened with reagent water
and dry thoroughly.
7.
Place the new roller cage on the shaft.
8.
Turn the roller cage on the shaft until the roller cage stops.
9.
Press the roller cage down until the cage snaps into place.
10. Reinstall the tubing as described in
Replacing the Pump Tubing
, page 3-39.
11. Press
Continue
.
A wash sequence starts. When the wash sequence finishes, a message box
appears prompting you to perform a two-point calibration.
12. Press
Yes
to perform a two-point calibration.