Bayer HealthCare Rapidlab 800 User Manual
Page 540

EĆ14
800 Series Operator's Manual
Table EĆ12.āă840 Recovery and Precision TestingĊpCO
2
Level
Mode
n
WRSD
*
Observed
Expected
Recovery
1
Syringe
24
0.08
14.0
14.3
98.0
Capillary
18
0.32
13.6
14.3
95.1
Microcapillary
24
0.34
14.0
14.3
97.6
Microsyringe
24
0.13
14.3
14.3
99.9
Expired Gas
24
0.10
14.5
14.3
101.4
2
Syringe
24
0.14
21.6
21.4
100.8
Capillary
18
0.17
21.4
21.4
100.0
Microcapillary
24
0.45
21.4
21.4
100.2
Microsyringe
24
0.15
21.6
21.4
101.2
Expired Gas
24
0.14
21.6
21.4
100.7
3
Syringe
24
0.22
35.8
35.7
100.4
Capillary
18
0.53
35.3
35.7
98.8
Microcapillary
23
0.77
36.2
35.7
101.3
Microsyringe
24
0.30
35.7
35.7
99.9
Expired Gas
24
0.25
35.5
35.7
99.4
4
Syringe
24
0.28
50.1
49.9
100.3
Capillary
18
0.40
51.3
49.9
102.8
Microcapillary
24
1.28
51.5
49.9
103.2
Microsyringe
24
0.21
50.3
49.9
100.8
Expired Gas
24
0.37
49.4
49.9
99.0
5
Syringe
24
0.26
71.5
71.3
100.3
(Continued)