Recovery and precision with whole blood – Bayer HealthCare Rapidlab 800 User Manual
Page 575

Performance Characteristics
EĆ49
Parameter
TotSD
[
WRSD
*
Mean
n
Level
FCOHb
1
96
52.64
0.18
0.35
2
96
32.03
0.16
0.30
3
96
9.52
0.35
0.39
FMetHb
1
96
5.05
0.13
0.23
2
96
4.70
0.14
0.24
3
96
4.07
0.19
0.30
FHHb
1
96
36.52
0.30
0.44
2
96
49.71
0.25
0.36
3
96
64.40
0.30
0.41
* WRSD = within-run standard deviation
[
TotSD = total standard deviation
Recovery and Precision with Whole Blood
For testing syringe and capillary modes, blood was collected in heparinized vacuum
tubes. It was tonometered at 37
°
C and/or adjusted chemically. Three levels of tHb
were prepared by separating the red cells from the plasma and recombining
appropriately. For each run, three replicates of each level were run on four 800
systems.
Table E-44 through Table E-46 summarize the results of the 800 CO-ox module
recovery and precision testing.
Table EĆ44.āă800 COĆox Module Recovery and Precision with Total Hemoglobin of
5-8 g/dL and Oxyhemoglobin > 80%
Analyte
Mode
n
Mean
Reference
Mean Difference
from Reference
WRSD
*
tHb
Capillary
72
6.64
–0.19
0.30
Syringe
66
6.53
–0.27
0.11
FO
2
Hb
Capillary
72
91.96
1.16
0.45
(Continued)