beautypg.com

Viewing and printing leveyćjennings charts – Bayer HealthCare Rapidlab 800 User Manual

Page 150

background image

2Ć52

Rapidlab 800 Operator's Manual

5.

Press

Home

to return to the Ready screen.

Viewing and Printing LeveyĆJennings Charts

Use this procedure to view and print Levey-Jennings charts of the current month’s
QC data.

You can print Levey-Jennings charts for the previous month by archiving the
previous month’s QC data and then printing Levey-Jennings charts of the archived
data. Refer to Archiving QC Data and Viewing Archived QC Data in Section 5.

1.

Access the Recall QC Data screen from the Menu screen:

a. Select

4 Data Recall

and press

Enter

.

b. Select

2 QC Data

and press

Enter

.

The QC Data Search Criteria screen appears.

2.

Type only the file number and press

Enter

.

The QC Data Search Criteria Log screen appears.

3.

Select

Reporting Options

.

NOTE:

You can view the previous month’s Levey-Jennings charts before you

analyze the first QC sample of the current month. After you analyze the first QC
sample of the current month, Levey-Jennings charts contain data only for the
current month.

4.

Select either Levey-Jennings Current Month or Levey-Jennings Previous
Month.

5.

Press

OK

.

6.

The Levey-Jennings Parameter Selection screen appears.

7.

Select the parameter you want to display on the chart and press

Enter

.

8.

Press

Done

.

The Levey-Jennings Chart appears, as shown in Figure 2-30.

4

Menu Code

2